作者
Jonatan Kutchinsky, Søren Friis, Margit Asmild, Rafael Taboryski, Simon Pedersen, Ras K Vestergaard, Rasmus B Jacobsen, Karen Krzywkowski, Rikke L Schrøder, Trine Ljungstrøm, Nathalie Hélix, Claus B Sørensen, Morten Bech, Niels J Willumsen
发表日期
2003/10/1
期刊
Assay and drug development technologies
卷号
1
期号
5
页码范围
685-693
出版商
Mary Ann Liebert, Inc.
简介
Planar silicon chips with 1-2-μm etched holes (average resistance: 2.04 ± 0.02 MΩ in physiological buffer, n = 274) have been developed for patch-clamp recordings of whole-cell currents from cells in suspension. An automated 16-channel parallel screening system, QPatch 16, has been developed using this technology. A single-channel prototype of the QPatch system was used for validation of the patch-clamp chip technology. We present here data on the quality of patch-clamp recordings and from actual drug screening studies of human potassium channels expressed in cultured cell lines. Using Chinese hamster ovary (CHO) and human embryonic kidney cells (HEK), gigaseals of 4.1 ± 0.4 GΩ (n = 146) and high-quality whole-cell current recordings were obtained from hERG and KCNQ4 potassium channels. Success rates for gigaseal recordings varied from 40 to 95%, and 67% of the whole …
引用总数
200420052006200720082009201020112012201320142015201620172018201920202021202220232024318310611527712573127131