Dan Wu, Jiangyang Zhang
发表日期
2016/4/15
期刊
Magnetic Resonance in Medicine
期号
doi: 10.1002/mrm.26242
简介
Purpose
We investigated the effect of microcirculatory flow on oscillating gradient spin echo (OGSE) diffusion MRI at low b‐values and developed a diffusion preparation method called diffusion encoding with dual‐frequency orthogonal gradients (DEFOG) to suppress the effect.
Methods
Compared to conventional OGSE sequences, DEFOG adds a pulsed gradient that is orthogonal to the oscillating gradient and has a moderate diffusion weighting (e.g., 300 s/mm2). In vivo MRI data were acquired from adult mouse brains (n = 5) on an 11.7 Tesla scanner, with diffusion times from 23.2 to 0.83 ms and b‐values from 50 to 700 s/mm2.
Results
Apparent diffusion coefficients (ADCs) measured using a conventional OGSE sequence at low b‐values (< 200 mm2/s) were significantly higher than those measured at moderate b‐values (> 300 mm2/s), potentially due to contributions from microcirculatory flow. In …