作者
Suraj D Serai, Jonathan R Dillman, Andrew T Trout
发表日期
2017/7
期刊
Radiology
卷号
284
期号
1
页码范围
244-254
出版商
Radiological Society of North America
简介
Purpose
To determine the agreement of proton density fat fraction (PDFF) measurements obtained with hepatic magnetic resonance (MR) imaging among readers, imager manufacturers, and field strengths.
Materials and Methods
This HIPAA-compliant study was approved by the institutional review board. After providing informed consent, 24 adult volunteers underwent imaging with one 1.5-T MR unit (Ingenia; Philips Healthcare, Best, the Netherlands) and two different 3.0-T units (750 W [GE Healthcare, Waukesha, Wis] and Ingenia) on the same day to estimate hepatic PDFF. A single-breath-hold multipoint Dixon-based acquisition was performed with commercially available pulse sequences provided by the MR imager manufacturers (mDIXON …
引用总数
2017201820192020202120222023202431915121016134