作者
Lennert De Ruiter, Dennis Janssen, Adam Briscoe, Nico Verdonschot
发表日期
2017/12
期刊
Journal of experimental orthopaedics
卷号
4
页码范围
1-8
出版商
Springer Berlin Heidelberg
简介
Background
Conventional total knee replacement designs show high success rates but in the long term, the stiff metal components may affect bone quality of the distal femur. In this study we introduce an all-polymer total knee replacement device containing a PEEK femoral component on an UHMWPE tibial implant and study its mechanical integrity, fixation, and stress shielding of the periprosthetic femur.
Methods
The implant was analysed in finite element simulations of level gait, adopted from the ISO 14243-1 standard. Mechanical integrity of the implant and underlying cement mantle were tested, and the fixation strength of the cement-implant interface was studied. Stress shielding was assessed based on strain energy density distributions in the distal femur. We compared PEEK and CoCr implants for mechanical performance and fixation, and …
引用总数
20172018201920202021202220232024411051067