The use of a Dynamic Headspace System (DHS) device combined with a Thermal Desorption Unit (TDU) interfaced to a Gas Chromatography/Mass Spectrometry (GC/MS) system is proposed for the determination of furfurals in oenological products. An experimental design protocol has been employed for the optimization of the instrumental settings concerning DHS and TDU extraction and desorption steps. It has been possible to individuate the following optimized conditions: incubation temperature 40 °C, purge volume 800 mL, dry volume 1500 mL, TDU hold time 5 min and incubation time 10 min. The performance of two different SPE sorbents, namely Tenax TA and Tenax GR used for the furfurals trapping, was investigated too.
The developed DHS sampling procedure showed good reproducibility values with a RSD% lower than 10% for all the monitored species. The optimized experimental settings have been …
引用总数
201220132014201520162017201820192020202120222023244334362221