John D Newell Jr, Matthew K Fuld, Thomas Allmendinger, Jered P Sieren, Kung-Sik Chan, Junfeng Guo, Eric A Hoffman
发表日期
2015/1/1
期刊
Investigative radiology
卷号
50
期号
1
页码范围
40-45
出版商
LWW
简介
Objectives
The purpose of this study was to evaluate the impact of ultralow radiation dose single-energy computed tomographic (CT) acquisitions with Sn prefiltration and third-generation iterative reconstruction on density-based quantitative measures of growing interest in phenotyping pulmonary disease.
Materials and Methods
The effects of both decreasing dose and different body habitus on the accuracy of the mean CT attenuation measurements and the level of image noise (SD) were evaluated using the COPDGene 2 test object, containing 8 different materials of interest ranging from air to acrylic and including various density foams. A third-generation dual-source multidetector CT scanner (Siemens SOMATOM FORCE; Siemens Healthcare AG, Erlangen, Germany) running advanced modeled iterative reconstruction (ADMIRE) software (Siemens Healthcare AG) was used.