简介
Background
Eye Movement Perimetry (EMP) uses Saccadic Eye Movement (SEM) responses for visual field evaluation. Previous studies have demonstrated significant delay in initiation of SEMs among glaucoma patients in comparison with healthy subjects. The aim of the current study was to develop an EMP-based screening grid to identify glaucomatous visual field defects.
Methods
An interactive test consisting of 36 locations and two stimulus contrasts (162 cd/m2 and 190 cd/m2 on a background of 140 cd/m2) was evaluated in 54 healthy subjects and 50 primary glaucoma patients. Each subject was presented a central fixation target combined with the random projection of Goldmann size III peripheral targets. Instructions were given to look at each peripheral target on detection and then re-fixate at the central fixation target while the saccades were …
引用总数
2019202020212022202322322
学术搜索中的文章
NS Kadavath Meethal, D Mazumdar, R Asokan… - Graefe's Archive for Clinical and Experimental …, 2018