作者
Yong-Chan Ha, Shin-Yoon Kim, Hee Joong Kim, Jeong Joon Yoo, Kyung-Hoi Koo
发表日期
2007/5/1
期刊
Clinical Orthopaedics and Related Research®
卷号
458
页码范围
106-110
出版商
LWW
简介
Advances in technology have reduced the risk of fracture of ceramic total hip arthroplasty implants, but concerns remain about fracture of both components. We retrospectively reviewed 133 patients (157 hips) who had cementless alumina-on-alumina total hip arthroplasties with a sandwich-type acetabular component. Six patients (seven hips) died and five patients (six hips) were interviewed by telephone (95% followup). The 122 patients (144 hips) examined had a minimum followup of 36 months (average, 45 months; range, 36-68 months). All acetabular cups and femoral stems were radiographically stable at the last followup. Five hips in five patients (3.5%) were revised because of ceramic liner fractures. Ceramic liner fractures occurred at a mean of 35 months (range, 24-48 months) postoperatively. Acetabular cups in the fracture group (n= 5) were more anteverted than those in the nonfracture group (n= 139 …
引用总数
20072008200920102011201220132014201520162017201820192020202120222023202410615141816181719101065117351
学术搜索中的文章
YC Ha, SY Kim, HJ Kim, JJ Yoo, KH Koo - Clinical Orthopaedics and Related Research®, 2007