作者
Benjamin Gilles, Rosalind Perrin, Nadia Magnenat-Thalmann, Jean-Paul Vallee
发表日期
2005/10/1
期刊
Academic radiology
卷号
12
期号
10
页码范围
1285-1292
出版商
Elsevier
简介
RATIONALE AND OBJECTIVES
For diagnosis, preoperative planning and postoperative guides, an accurate estimate of joint kinematics is required. It is important to acquire joint motion actively with real-time protocols.
MATERIALS AND METHODS
We bring together MRI developments and new image processing methods in order to automatically extract active bone kinematics from multi-slice real-time dynamic MRI. We introduce a tracking algorithm based on 2D/3D registration and a procedure to validate the technique by using both dynamic and sequential MRI, providing a gold standard bone position measurement.
RESULTS
We present our technique for optimizing jointly the tracking method and the acquisition protocol to overcome the trade-off in acquisition time and tracking accuracy. As a case study, we apply this methodology on a human hip joint.
CONCLUSION
The final protocol (bFFE, TR/TE 3.5/1.1 ms …