Weitian Chen, Atsushi Takahashi, Eric Han
发表日期
2011/6/1
期刊
Magnetic resonance imaging
卷号
29
期号
5
页码范围
608-619
出版商
Elsevier
简介
T1ρ imaging is useful in a number of clinical applications. T1ρ preparation methods, however, are sensitive to non-uniformities of the B0 magnetic field and the B1 RF field. These common system imperfections can result in image artifacts and quantification errors in T1ρ imaging. We report on a phase-cycling method which can eliminate B1 RF inhomogeneity effects in T1ρ imaging. This method does not only correct for image artifacts but also for T2ρ contamination caused by B1 RF inhomogeneity. The presence of B0 magnetic field inhomogeneity can compromise the effectiveness of this method for B1 RF inhomogeneity correction. We demonstrate that, by combining the spin-locking scheme reported by Dixon et al. (Myocardial suppression in vivo by spin locking with composite pulses. Magn Reson Med 1996; 36:90–94) with phase cycling, we can simultaneously correct B0 magnetic field inhomogeneity effects and …