作者
J Tibballs, M Hochmann, A Osborne, B Carter
发表日期
1992/8
期刊
Anaesthesia and intensive care
卷号
20
期号
3
页码范围
326-331
出版商
SAGE Publications
简介
Changes in thoracic electrical bioimpedance during the cardiac cycle are utilised by the BoMed NCCOM3 monitor to measure cardiac output (COTEB). The technique provides a continuous noninvasive measurement but it has not been widely accepted. To determine the accuracy of the monitor, we compared its measurement with cardiac output measured by dye dilution (COdd) during induced hypotension and recovery in 23 dogs. After calibration of the NCCOM3 monitor during a resting state in each dog [mean blood pressure 112 ± 17 (SD), mean CODD 3.22 ± 0.99 l/min], the mean difference (COTEB-CODD) between paired measurements at the nadir of hypotension (bloodpressure 55 ± 24 mmHg) was 0.29 ± 0.47 l/min whose limits of agreement (mean difference ± 2 SD) were + 111.8% and -59.1% of the mean hypotensive COdd (110 ± 0.66 l/min). Upon recovery from hypotension (mean blood pressue 102 ± …