Piotr Zygmanski, Florin Rosca, Dnyanesh Kadam, Friedlieb Lorenz, Adrian Nalichowski, Laurence Court, Lee Chin
发表日期
2007/9
期刊
Journal of Applied Clinical Medical Physics
卷号
8
期号
4
页码范围
76-95
简介
Intensity‐modulated radiation therapy (IMRT) plans for the treatment of large and complex volumes may contain a relatively large contribution from multileaf collimator (MLC) transmission. In such cases, comprehensive characterization of direct and scatter MLC transmission is important. We designed a set of tests (open beam, closed static MLC, and dynamic MLC gap) to determine dosimetric MLC properties as a function of field size and depth at the central axis.
We developed a generalized model of MLC transmission to account for direct MLC transmission, MLC scatter, beam hardening, and leaf‐end transmission (dosimetric gap). The model is consistent with the beam model used in IMRT optimization. We tested the model for extreme asymmetric fields relevant for large targets and for split IMRT fields. We applied our MLC scatter estimation formula to clinically relevant cases and showed that MLC scatter is …