Giacomo Pucci, Joseph Cheriyan, Annette Hubsch, Stacey S Hickson, Parag R Gajendragadkar, Timothy Watson, Michael O'Sullivan, Jean Woodcock-Smith, Giuseppe Schillaci, Ian B Wilkinson, Carmel M McEniery
发表日期
2013/1/1
期刊
Journal of hypertension
卷号
31
期号
1
页码范围
77-85
出版商
LWW
简介
Objectives:
The Vicorder is a new brachial cuff-based device that estimates central blood pressure (cBP) using a brachial-to-aortic transfer function. The aim of this study was to evaluate cBP estimated by the Vicorder.
Methods:
During cardiac angiography, cBP estimated by the Vicorder and the SphygmoCor was evaluated against simultaneous invasive cBP in 50 patients. The two devices were also compared noninvasively in a separate group of 90 healthy individuals.
Results:
Central SBP (cSBP) obtained with each device satisfied the American Association for the Advancement of Medical Instrumentation accuracy criteria when peripheral waveforms were calibrated to invasive mean arterial pressure (MAP)/DBP: estimated− invasive cSBP difference,− 4.0±7.4 mmHg, Vicorder, P< 0.001;− 1.4±7.9 mmHg, SphygmoCor, P= 0.21. When oscillometric brachial SBP/DBP was used for peripheral waveform calibration, cSBP …