作者
Matthew M Pier, Luke J Pasick, Daniel A Benito, Prashant Saini, Sahil Patel, Arjun S Joshi, Joseph F Goodman, Punam Thakkar
发表日期
2021/5
期刊
Journal of Reconstructive Microsurgery
卷号
37
期号
04
页码范围
365-371
出版商
Thieme Medical Publishers, Inc.
简介
Background Implantable Dopplers (IDs) are widely used for postoperative free flap vascular monitoring. However, IDs may contribute to free flap complications or failure and better understanding of device malfunctions is needed.
Methods The U.S. Food and Drug Administration's Manufacturer and User Facility Device Experience database was queried for all reports of ID adverse events from two leading manufacturers (Cook Vascular and Synovis Life Technologies) in free flap surgery from January 2010 to March 2020. Reports were reviewed and categorized. A comparison of reoperations within select categories was performed using Chi-square analysis.
Results Of 209 included reports, the most common device malfunctions were venous anastomotic coupler misalignment (35.4%) and coupler ring detachment (24.4%). Synovis devices were used in 100% of reports of vessel compression and Cook …
引用总数
学术搜索中的文章
MM Pier, LJ Pasick, DA Benito, P Saini, S Patel… - Journal of Reconstructive Microsurgery, 2021