作者
Bernhard A Herzog, Ronny R Buechel, Lars Husmann, Aju P Pazhenkottil, Irene A Burger, Mathias Wolfrum, Rene N Nkoulou, Ines Valenta, Jelena R Ghadri, Valerie Treyer, Philipp A Kaufmann
发表日期
2010/10/1
期刊
Journal of Nuclear Medicine
卷号
51
期号
10
页码范围
1539-1544
出版商
Society of Nuclear Medicine
简介
The aim of this study was to validate attenuation correction (AC) using low-dose standard CT for myocardial perfusion imaging (MPI) on a novel ultra fast γ-camera with cadmium-zinc-telluride (CZT) detector technology. Methods: Sixty-six patients (body mass index ± SD, 27.2 ± 3.5 kg/m2; range, 19.1–36.0 kg/m2) underwent a 1-d 99mTc-tetrofosmin adenosine stress–rest imaging protocol with 15-min acquisitions on a standard dual-head SPECT camera. All scans were repeated within minutes on the CZT camera, with 3-min acquisitions for stress (low dose) and 2-min acquisitions for rest (high dose) as recently established. We compared maximum myocardial uptake (20-segment model) from CZT versus standard SPECT MPI by intraclass correlation without and with CT AC. In addition, clinical agreement for each coronary territory for all scans from both devices was assessed, and Bland–Altmann (BA) limits of …
引用总数
20112012201320142015201620172018201920202021202220234967993562522