Validation of Watch-PAT-200 against polysomnography during pregnancy

LM O'Brien, AS Bullough, AV Shelgikar… - Journal of Clinical …, 2012 - jcsm.aasm.org
LM O'Brien, AS Bullough, AV Shelgikar, MC Chames, R Armitage, RD Chervin
Journal of Clinical Sleep Medicine, 2012jcsm.aasm.org
Study Objectives: To determine the relationships between key variables obtained from
ambulatory polysomnography (PSG) and the wrist-worn Watch-PAT 200 device in pregnant
women. Methods: In this prospective cohort study, women in their third trimester of
pregnancy underwent full overnight home PSG using the 22-channel MediPalm system and
the Watch-PAT 200 device. PSGs were scored by a blinded, experienced technologist using
AASM 2007 criteria; the Watch-PAT was scored automatically by the manufacturer's …
Study Objectives
To determine the relationships between key variables obtained from ambulatory polysomnography (PSG) and the wrist-worn Watch-PAT 200 device in pregnant women.
Methods
In this prospective cohort study, women in their third trimester of pregnancy underwent full overnight home PSG using the 22-channel MediPalm system and the Watch-PAT 200 device. PSGs were scored by a blinded, experienced technologist using AASM 2007 criteria; the Watch-PAT was scored automatically by the manufacturer's proprietary software.
Results
A total of 31 pregnant women were studied. Mean age was 30.2 ± 7.1 years; mean gestational age was 33.4 ± 3.0 weeks; mean BMI was 31.9 ± 8.1 kg/m2; 39% of women were nulliparous. Key variables generated by PSG and Watch-PAT correlated well over a wide range, including the apnea-hypopnea index (AHI, r = 0.76, p < 0.001); respiratory disturbance index (RDI, r = 0.68, p < 0.001), mean oxygen saturation (r = 0.94, p < 0.001), and minimum oxygen saturation (r = 0.88, p < 0.001). The area under the curve for AHI ≥ 5 and RDI ≥ 10 were 0.96 and 0.94, respectively. Association between stage 3 sleep on PSG and deep sleep on Watch-PAT was poor. Watch-PAT tended to overscore RDI, particularly as severity increased.
Conclusions
Among pregnant women, Watch-PAT demonstrates excellent sensitivity and specificity for identification of obstructive sleep apnea, defined as AHI ≥ 5 on full PSG. Watch-PAT may overestimate RDI somewhat, especially at high RDI values.
Citation
O'Brien LM; Bullough AS; Shelgikar AV; Chames MC; Armitage R; Chervin RD. Validation of Watch-Pat-200 against polysomnography during pregnancy. J Clin Sleep Med 2012;8(3):287-294.
American Academy of Sleep Medicine