Feasibility of late acquisition [68Ga] Ga-PSMA-11 PET/CT using a long axial field-of-view PET/CT scanner for the diagnosis of recurrent prostate cancer—first clinical …

I Alberts, G Prenosil, C Mingels, KP Bohn… - European journal of …, 2021 - Springer
I Alberts, G Prenosil, C Mingels, KP Bohn, M Viscione, H Sari, A Rominger
European journal of nuclear medicine and molecular imaging, 2021Springer
Purpose While acquisition of images in [68 Ga] Ga-PSMA-11 following longer uptake times
can improve lesion uptake and contrast, resultant imaging quality and count statistics are
limited by the isotope's half-life (68 min). Here, we present a series of cases demonstrating
that when performed using a long axial field-of-view (LAFOV) PET/CT system, late imaging
is feasible and can even provide improved image quality compared to regular acquisitions.
Methods In this retrospective case series, we report our initial experiences with 10 patients …
Purpose
While acquisition of images in [68 Ga]Ga-PSMA-11 following longer uptake times can improve lesion uptake and contrast, resultant imaging quality and count statistics are limited by the isotope’s half-life (68 min). Here, we present a series of cases demonstrating that when performed using a long axial field-of-view (LAFOV) PET/CT system, late imaging is feasible and can even provide improved image quality compared to regular acquisitions.
Methods
In this retrospective case series, we report our initial experiences with 10 patients who underwent standard imaging at 1 h p.i. following administration of 192 ± 36 MBq [68 Ga]Ga-PSMA-11 with additional late imaging performed at 4 h p.i. Images were acquired in a single bed position for 6 min at 1 h p.i. and 16 min p.i. at 4 h p.i. using a LAFOV scanner (106 cm axial FOV). Two experienced nuclear medicine physicians reviewed all scans in consensus and evaluated overall image quality (5-point Likert scale), lesion uptake in terms of standardised uptake values (SUV), tumour to background ratio (TBR) and target-lesion signal to background noise (SNR).
Results
Subjective image quality as rated on a 5-point Likert scale was only modestly lower for late acquisitions (4.2/5 at 4 h p.i.; 5/5 1 h p.i.), TBR was significantly improved (4 h: 3.41 vs 1 h: 1.93, p < 0.001) and SNR was improved with borderline significance (4 h: 33.02 vs 1 h: 24.80, p = 0.062) at later imaging. Images were obtained with total acquisition times comparable to routine examinations on standard axial FOV scanners.
Conclusion
Late acquisition in tandem with a LAFOV PET/CT resulted in improvements in TBR and SNR and was associated with only modest impairment in subjective visual imaging quality. These data show that later acquisition times for [68 Ga]Ga-PSMA-11 may be preferable when performed on LAFOV systems.