Measurement of Urinary Cystatin C with a Particle‐Enhanced Turbidimetric Immunoassay on Architect C i8200
FH Noraddin, M Flodin, A Fredricsson… - Journal of Clinical …, 2012 - Wiley Online Library
… Urinary cystatin C measurements were performed on Abbott Architect ci8200 analyzer (Abbott
Park, … The immune complexes formed by agglutination were measured at 444 nm and the …
Park, … The immune complexes formed by agglutination were measured at 444 nm and the …
Measurement and validation of urinary cystatin C by particle-enhanced turbidimetric immunoassay on Architect ci8200
F Hikmet Noraddin - 2011 - diva-portal.org
… These finding may indicate urinary cystatin C as a broad marker for tubular damage in … of
urinary analysis of cystatin C was evaluated with particle enhanced turbidimetric immunoassay (…
urinary analysis of cystatin C was evaluated with particle enhanced turbidimetric immunoassay (…
Particle enhanced turbidimetric immunoassay for the determination of urine cystatin C on Cobas c501
A Sohrabian, FH Noraddin, M Flodin, A Fredricsson… - Clinical …, 2012 - Elsevier
… measured urinary cystatin C concentration versus the theoretically expected values. The
highest calibrator point used had a concentration of 1.09 mg/L. None of the absorbance values …
highest calibrator point used had a concentration of 1.09 mg/L. None of the absorbance values …
Novel automated immune turbidimetric assay for routine urinary cystatin-C determinations
B Szirmay, P Kustan, Z Horváth-Szalai, A Ludány… - Bioanalysis, 2018 - Future Science
… For u-CYSC method adaptation, validation and measurements a particle-enhanced turbidimetric
assay … In our work, a novel automated particle enhanced immune turbidimetric assay for …
assay … In our work, a novel automated particle enhanced immune turbidimetric assay for …
Performance evaluation of a turbidimetric cystatin C assay on different high-throughput platforms
…, V Stove, H Luthe, KH Rhode, C Beck… - … journal of clinical and …, 2010 - Taylor & Francis
… particle-enhanced immunoturbidimetry assay of cystatin C from Gentian (Moss, Norway) was
used on both Abbott Architect ci8200 (… Besides that, plasma cystatin C measurements were …
used on both Abbott Architect ci8200 (… Besides that, plasma cystatin C measurements were …
Performance characteristics of a cystatin C immunoassay with avian antibodies
K Sunde, T Nilsen, M Flodin - Upsala Journal of Medical Sciences, 2007 - Taylor & Francis
… , measured over 20 days with two lots, was =4.2%. Comparison with a particle enhanced
nephelometric cystatin C immunoassay (… cystatin C immunoassay between Architect ci8200 and …
nephelometric cystatin C immunoassay (… cystatin C immunoassay between Architect ci8200 and …
Performance evaluation of a particle-enhanced turbidimetric cystatin C assay using the Abbott Aeroset analyser and assessment of cystatin C-based equations for …
Y Sun, T Jiang, Z Zeng, P Chen - Nephrology Dialysis …, 2010 - academic.oup.com
… Ltd, Shanghai, China) on Architect Aeroset analyzer (Abbott … 917 analyser [32] and the
Abbott ci8200 analyzer [33], have reported … , measurement of cystatin C using a particle-enhanced …
Abbott ci8200 analyzer [33], have reported … , measurement of cystatin C using a particle-enhanced …
Cystatin C concentration during the first three postnatal days in healthy term newborns
… Cys-C was measured within first three days of birth using particle-enhanced turbidimetric
immunoassay (PETIA) on the Architect plus ci8200 analyzer. …
immunoassay (PETIA) on the Architect plus ci8200 analyzer. …
Analytical performances of PENIA and PETIA urinary cystatin C determination allow tubular injury investigation
AS Bargnoux, V Azoury, S Badiou… - Annals of Clinical …, 2019 - journals.sagepub.com
… CysC in urine was measured by a particle-enhanced nephelometric immunoassay using
Siemens reagents and by a particle-enhanced turbidimetric immunoassay using DiaSys …
Siemens reagents and by a particle-enhanced turbidimetric immunoassay using DiaSys …
[PDF][PDF] Agreement between creatinine and cystatin C-based equations in the estimation of glomerular filtration rate among Malaysian patients with renal impairment
… the agreement of eGFR measured by using cystatin C- and … picrate method on an
Architect C System instruments (Abbott … Serum cystatin C levels were assessed by particleenhanced …
Architect C System instruments (Abbott … Serum cystatin C levels were assessed by particleenhanced …