The cementless artificial hip cup--implant components with reliable long-term stability?

D Gebauer, M Schug - Biomedizinische Technik. Biomedical …, 1990 - europepmc.org
D Gebauer, M Schug
Biomedizinische Technik. Biomedical Engineering, 1990europepmc.org
A hip joint simulator was developed to analyse the mechanism of loosening of cementless
artificial hip cups. The machine induces vibrating motions and asymmetrical tilt shock
loadings of the artificial cup. On measuring the primary stability of threaded cups, the
simulation tests performed on pelvis substitute models, and animal and human acetabula
failed to show any loosening of threaded cups fixed in place with a screw-in torque of more
than 10 Nm. Instable cups became loose all the sooner, the lower the preload between the …
A hip joint simulator was developed to analyse the mechanism of loosening of cementless artificial hip cups. The machine induces vibrating motions and asymmetrical tilt shock loadings of the artificial cup. On measuring the primary stability of threaded cups, the simulation tests performed on pelvis substitute models, and animal and human acetabula failed to show any loosening of threaded cups fixed in place with a screw-in torque of more than 10 Nm. Instable cups became loose all the sooner, the lower the preload between the cup and femoral head. This demonstrates the importance of both the screw-in torque of the cup and the training status of the periarticular muscles.